Monitoring
Recording
3-10
Record on Shock
Record on Alarms
The automatic recordings for both delayed and non-delayed recorder modes of
operation are defined in Table 3-7, “Automatic Recordings,” on page 3-10.
Post Shock Data
As described in Appendix A, you can enable or disable the recording of post shock
statistics in setup menu 2.
If Post Shock Data is enabled, the defibrillator will record the shock delivery sta-
tistics (Actual Delivered Energy, Patient Impedance, Peak Current).
If Post Shock Data is disabled, the defibrillator will record the energy to which it
was charged as the delivered energy. For example, if the unit was charged to 200J,
the delivered energy annotation on the ECG strip would be DEL 200J.
Table 3-7 Automatic Recordings
Event
Delayed mode
Pre-event time
Delayed mode
Post-event
time
Non- Delayed mode
Post-event time
pressed 6 seconds 3 seconds 3 seconds
Charge 6 seconds Until Shock or
Disarm event
Until Shock or Disarm
event.
Shocking the
patient
6 seconds 12 seconds 12 seconds
Alarms violation 6 seconds 3 seconds 6 seconds
Disarm 6 seconds 3 seconds 3 seconds
Test discharge N/A 3 seconds 3 seconds