Troubleshooting
Operational Checks
7-10
Keep hands clear of the paddle electrode edges. Use your thumbs
to depress the Shock buttons on the paddle handles.
3 With the paddles in their holders, grasp the paddle handles and press the Apex
paddle Shock button. Verify that the defibrillator does not discharge.
4 Release the Apex paddle Shock button, then press the Sternum paddle Shock
button. Verify that the defibrillator does not discharge.
5 Press to place the defibrillator in sync mode.
6 Press and hold both Shock buttons. Verify that the defibrillator does not dis-
charge.
7 Press again to remove the defibrillator from sync mode.
8 With the paddles in their holders, press and briefly hold both Shock buttons at
once.
9 The recorder will print a test report.
To check the instrument with the pads adaptor cable, perform the following steps.
1 Connect the Test Load (M1781A) to the Pads Adaptor Cable (M1750A/B).
2 Turn Energy Select to 100 joules.
3 Press Charge. Wait for the Charge Done indicators.
4 Press each Shock button independently. Verify that the defibrillator does not dis-
charge.
5 While depressing Charge, press and briefly hold both Shock buttons.
6 The recorder will print a test report.