Installation and Setup
Setup
A-8
1 Depress while turning the Energy Select control from Off
(Standby) to Monitor On. A menu will appear with the following choices.
CALIBRATE DEFIB
SETUP MENU 1
SETUP MENU 2
RESTORE FACTORY SETTINGS
PRINT LOG
TEST DEFIB
TEST ECG
TEST CRT
TEST RECORDER
TEST CONTROLS
TEST INDICATORS
TEST BATTERY
TEST PACER
2 Select SETUP MENU 1 by pressing until the highlight appears
on .
3 Press . SETUP MENU 1 will appear with the current setup values
displayed.
4 Press or until the highlight appears on the value
you wish to change.
5 Press .
6 Press or to scroll through the available choices for
this parameter.
7 When the choice you want is displayed, press to set your choice.
8 Press or until the highlight appears on the next
value you wish to change.
9 Repeat steps 5 through 8 until you are finished configuring the settings in setup
menu 1.
10 Depress both sides of the key at once to return to the setup/diag-
nostic menu.
Sync
HR
ECG Size
SETUP MENU 1
Lead
ECG Size
ECG Size
Lead
ECG Size
ECG Size
Lead
ECG Size
ECG Size
ECG Size