EasyManua.ls Logo

HP CodeMaster XL+ M1722B - Monitor

HP CodeMaster XL+ M1722B
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation and Setup
Specifications
A-12
Environmental Storage Conditions: -20 to 70 deg C, 90% relative humidity for 24
hours at 65 deg C, 15,000 ft altitude.
Monitor
Inputs: ECG may be viewed through paddles or patient cable. Lead I, II, III, or
PADDLES selectable. Additional leads (avR, avF, AvL, V Leads) and PADS are
available. Monitor and recorder indicate selected ECG source.
Lead Fault: LEADS OFF message and dashed baseline appear on monitor if a lead
becomes disconnected.
Common Mode Rejection: Greater than 100 dB measured as per AAMI standards
for cardiac monitors (EC13).
Display Size and Type: 5 inch (12.7 cm) diagonal CRT for 4 seconds of ECG data
on screen; non-fade, fixed trace. Scrolling trace is selectable.
Sweep Speed: 25 mm/sec nominal.
Frequency Response: 0.5 to 40 Hz.
Heart Rate Display: Digital readout on monitor from 15 to 300 BPM.
Heart Rate Alarms: Three configurable pairs of high and low heart rate alarm limits
from 20 to 280 BPM.
ECG Output: 1V/mV.
Patient Cable Length: 10 ft (3 m).
Thermal Array Recorder
Event Summary: Stores and prints 3 seconds pre- and 8 seconds post-critical event
data for up to 28 events. Data retained after unit is turned off.
Annotates: Time, date, HR, event marker, ECG mode, defibrillator mode, selected
energy, actual delivered energy, peak current, and patient impedance.
Speed: 25 mm/sec.

Table of Contents

Related product manuals