EasyManua.ls Logo

HP E1440A - Page 178

HP E1440A
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
+200,CbCommand not allowed; Cannot accept command
in
mode
SWEEPJLIST"
Triggered by: all commands that attempt to alter parameters in the
SWEEPILIST subsystem. Also *SAV, *TST? and
CAL?.
Switch
FREQ:MODE to CW or FIX beforehand.
+200,"Command not allowed; Cannot accept fiequency for DC function"
Triggered by :FREQ command. Attempt has been made to
enter frequency while output function is DC, the two actions are
incompatible
+2OO,"Command not allowed; Cannot accept phase for
D
C function"
Triggered by :PHAS command. Attempt to enter phase while output
function is DC, this is not possible
+2OO,"Command not allowed; Parameter is fixed for
TTL
function"
Triggered by :VOLT or V0LT:OFFS command. The main output is
always off in TTL function, the only output is the SYNC output.
The SYNC signal characteristic is not alter able.
-213,"Init ignored; Sweep is currently running or frequency mode is
FIX"
Triggered by :INIT. Attempt to start a sweep while sweep is already
running or while frequency mode is fixed (non sweeping).
-221,"Settings codict; AC Voltage in Vrms/dBm out of range in new
function"
-
Triggered by :FUNC command. The instrument maintains the
output voltage over function changes in the current default unit. If
the current default unit is not
V
(VRMS or dBm)
a
value legal for
one output function, may not be legal for another.
In
other words
:
the Vpp voltage that represents
X
dBm in SIN
function is different from the Vpp voltage for
X
dBm in SQU
function
.
-221,"Settings confiict; Active marker in logarithmic interval"
Triggered by :FREQ:MODE command. One of the in-sequence
intervals used, has spacing LOG and a marker ON.
-221,"Settings confiict; Active marker outside sweep span"
Triggered by :FREQ:MODE command. There is a marker switched
on outside the frequency span defined
by
start and stop.
-221,"Settings codict; Already another marker trigger on"
Triggered by :OUTP:TTLT:STAT and 0UTP:TTLT:SOUR
command. Cannot accept marker trigger on, because another marker
trigger is on. The marker trigger can only be applied to one line.
-221,"Settings confiict
;
Already another sync trigger on"
Triggered by :OUTP:TTLT:STAT and 0UTP:TTLT:SOUR
command. Cannot accept sync trigger on, because another sync
trigger is on. The sync trigger can only be applied to one line.