-221,"Settings codict;
AM
allowed for function sines only"
Triggered
by
:FUNC
or
AM
command.
It
is not permissible to switch
AM
on while function is not
SIN
and (vice versa) nor is it permissible
to select a function other than
SIN
while
AM
is on.
-221,"Settings confiict; Current output function is not sweep able"
Triggered by :FREQ:MODE command. Attempt to switch frequency
mode to SWEEP or LIST while function is DC or TTL.
-221,"Settings conflict; Frequency too high for post amplifier"
Triggered by :FREQ or 0UTP:AMPL command.
If
post amplifier
state ON, the frequency is limited to
1
MHz.
-221,"Settings confiict; Frequency too high for TTL trigger lines"
Triggered by :FREQ or 0UTP:TTLT command. If SYNC signal
is routed to one of the
8
TTL trigger lines and the line driver is
enabled, the frequency is limited to
10
MHz (VXI back plane
bandwidth).
-221,"Settings conflict; Lists are too short for sequence"
Triggered by: FREQ:MODE command. There is an index in the
sweep sequence that is higher than the length of the lists.
-221,"Settings conflict; List lengths are different and not 1"
Triggered by: FREQ:MODE command. All lists (start, stop, mark,
spacing, marker on and dwell) must have the same length or must
have the length
1.
(A list with length
1
is treated as a list of the
required length with the entry in slot
1
duplicated as often as
necessary.)
-221,"Settings conflict; No dBm allowed when post amplifier is active"
Triggered by: VOLT or 0UTP:AMPL command. It is not
permissible to program a dBm voltage while the post amplifier is
active and (vice versa) and it is not permissible to switch the post
amplifier on while the current voltage unit is
dBm.
-221,"Settings conflict; No room for markers"
Triggered by: FREQ:MODE command. Time distance between start
and stop frequency is too short to place a marker in between. (The
minimal distance between start
-b
marker, marker
-b
marker or
marker
-b
stop is
1.5
ms.)
-221,"Settings confiict; Span for LOG interval too low
(<
1 decade)"
Triggered by: FREQ:MODE command. Stop frequency of a LOG
interval is less than
10
times start frequency.
-221,"Settings conflict; Start fiequency for
LOG
interval too low
(<
1)"
Triggered by: FREQ:MODE command. Start frequency of one of the
LOG intervals is less than
1
Hz.
-221,"Settings conflict; Start frequency too high for post amplifier"
Triggered by: FREQ:MODE command.
If
post amplifier state ON,
the start frequency is limited to
1
MHz.
Error Messages
E-3