Verify product is set to match available line voltage before operation.
Minimize shock hazard by connecting chassis and cabinet to an electrical ground.
Avoid operating the instrument in the presence of flammable gases or fumes.
Overview of the E3610A, E3611A, E3612A dual range bench power supplies and their features.
Inspect the power supply for any obvious damage that may have occurred during shipment.
Proper placement and ensuring adequate airflow for cooling the instrument during operation.
Details on the power sources the supply can be operated from, based on the voltage option.
Information regarding the three-conductor power cable and the importance of cabinet ground connection.
Step-by-step guide to verify the power supply is operational using front panel controls and indicators.
Procedure for setting up the power supply for constant voltage output mode.
Procedure for setting up the power supply for constant current output mode.
Guidance on connecting loads to the output terminals to minimize coupling effects and noise.
Information on adjusting voltage or current above rated values and its effect on performance.
Explanation of how the power supply circuits function, including feedback loops and error amplifiers.
Procedures for verifying the unit's compliance with specifications and characteristics.
Measures the change in DC output voltage when load resistance changes from open to full load.
Measures the change in DC output voltage when AC input voltage changes within its rated range.
Procedure to measure the root-mean-square output voltage ripple and noise.
Measures the change in DC output current when load resistance changes from short to full load.
Measures the change in DC output current when AC input voltage changes within its rated range.
Procedure to measure the root-mean-square output current ripple and noise.
Procedure to measure the peak-to-peak output current ripple and noise.
Steps to calibrate ammeter and full scale current in the high current range.
Steps to calibrate ammeter and full scale current in the low current range.
Procedure to calibrate the voltage display to match the DVM reading.