6.1.1. Safety standards
Instrument safety: IEC/EN61010-1, IEC/EN61557-1
EMC : IEC/EN61326-1
Technical literature: IEC/EN61187
Accessory safety: IEC/EN61010-031, IEC/EN61010-2-032
Insulation: double insulation
Pollution degree: 2
Maximum height: 2000m (6561ft)
Measurement category: CAT III 240V to earth, maximum 415V between inputs
Earth resistance : IEC/EN61557-5
6.1.2. General features
Mechanical features
Dimensions (L x W x H): 240x100x45mm ; (9x4x2inches)
Weight (including batteries): 630g (22 ounces)
Power supply
Battery type: 4x1.5 V alkaline batteries type AA LR6
Low battery indication: symbol Errore. Non si possono creare oggetti
dalla modifica di codici di campo. is displayed
Battery life: approx. 500 measurements
Auto Power OFF: after 5 minutes of idleness
Display:
LCD 53mm x 53mm ; (2x2inches)
6.2. ENVIRONMENT
6.2.1. Environmental conditions
Reference temperature: 23°C ± 5°C ; (73°F ± 41°F)
Operating temperature: 0 ÷ 40 °C ; (32°F ÷ 104 °F)
Relative humidity: <80%RH
Storage temperature: -10 ÷ 60 °C ; (14°F ÷ 140 °F)
Storage humidity: <80%RH
This instrument satisfies the requirements of Low Voltage Directive 2014/35/EU
(LVD) and of EMC Directive 2014/30/EU
This instrument complies the requirements of 2011/65/EU (RoHS) directive and
2012/19/EU (WEEE) directive
6.3. ACCESSORIES