4.3.4. Diode test
CAUTION
Before attempting any resistance measurement, cut off power supply from
the circuit to be measured and make sure that all capacitors are discharged,
if present.
Fig. 5: Use of the instrument for diode test
1. Select position CAP.
2. Press the MODE key until the symbol is displayed.
3. Insert the red cable into input terminal VHz%CAPTemp and the black cable
into input terminal COM.
4. Position the leads at the ends of the diode to be tested, respecting the indicated
polarity (see Fig. 5). The value of directly polarized threshold voltage is shown on the
display.
5. If threshold value is equal to 0mV, the P-N junction of the diode is short-circuited.
6. If the display shows the message
O.L., the terminals of the diode are reversed with
respect to the indication given in Fig. 5 or the P-N junction of the diode is damaged.