Immunity Test IEC 60601 Test
Level
Compliance Level Electromagnetic
Environment—Guidance
Conducted
RF IEC
61000-4-6
Radiated RF IEC
61000-4-3
3
Vrms 150 kHz
to 80 MHz
3 V/m 80 MHz to
2.7 GHz
(3) Vrms
(
3) V/m
Portable and mobile communications
equipment should be separated from the T7
CART by no less than the distances calculated/
listed below:
D=(3.5/V1)(Sqrt P)
D=(12/V2)(Sqrt P)
D=(12/E1)(Sqrt P)
80 to 800 MHz where P is the max power in
watts and D is the recommended separation
distance in meters. Field strengths from
fixed transmitters, as determined by an
electromagnetic site survey, should be less
than the compliance levels (V1 and E1).
Interference may occur in the vicinity of
equipment containing a transmitter.
D=(23/E1)(Sqrt P)
800 MHz to 2.7 GHz where P is the max
power in watts and D is the recommended
separation distance in meters. Field strengths
from fixed transmitters, as determined by an
electromagnetic site survey, should be less than
the compliance levels (V1 and E1). Interference
may occur in the vicinity of equipment
containing a transmitter.
Equipment and Systems that are
not Life-Supporting
Do not dispose of the T7 or any associated components at the end of service life. Contact Humanscale
for disposal instructions.