Product Specification
FM800Encore Service Manual 10
4.9 Maternal blood pressure
Method Oscillometric
Measurement ranges
Systolic 50–280 mmHg Diastolic 10–260 mmHg Pulse 40–240
bpm
Accuracy ±2% or 3mmHg, whichever is greater
Modes
Manual or automatic User-selectable interval in Auto Mode: 3, 5,
10, 15, 20, 30, 45, 60, 90 or 120 minutes
Record/display
On-screen display and printed record of:
Systolic blood pressure.
Diastolic blood pressure.
Pulse rate.
Printed record of:
Mean arterial pressure.
[Note: Mean arterial pressure not shown in the USA or Canada]
Alarms
Systolic High: 55–255 mmHg Systolic Low: 50–250 mmHg
Diastolic High: 35–220 mmHg Diastolic Low: 30–215 mmHg
System: air leaks. movement, overpressure etc
Safety
Type CF protection Hardware and software controls to limit:
Inflation (max. 300 mmHg)
Measurement time (max. 160 secs)
Standards ANSI/AAMI SP10-1992
4.10 Maternal oximetry
Sensor types
(a) The following probe is suitable for FM830Encore equipped
with BCI-B1 MSpO2 modules:
Smiths Medical 3444 singer sensor
Do not use any sensors other than these.
(b) The following probes are suitable for use with the
FM830Encore models equipped with BCI-B2 Maternal
pulse oximetry technology:-
3444 Finger Sensor
Nellcor
TM
Durasensor
®
DS-100A
Do not use the oximetry sensors during magnetic
resonance imaging (MRI) scanning. Induced current could cause
burns. The oximeter may affect the MRI image, and the MRI unit
may affect the accuracy of oximetry measurements.
Saturation range 1-99% SpO2
Saturation accuracy
±2 digits (70-100%)
Less than 70% is undefined.
Pulse rate range 30-254 bpm
Pulse rate accuracy ±3 bpm
Record/display
On-screen display and printed record of:
Maternal % SpO2
Heart rate
Alarms
High and low saturation: 0-100% SpO2
Signal loss: Pulse or ECG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Huntleigh FM800Encore and is the answer not in the manual?

Huntleigh FM800Encore Specifications

General IconGeneral
TypeFetal/Maternal Monitor
DisplayLCD
Fetal Heart Rate Range50-210 BPM
Maternal Heart Rate Range30-240 BPM
Ultrasound Frequency1 MHz
Paper Speed1, 2, 3 cm/min
Power Supply100 - 240 VAC, 50/60 Hz
TOCO Range0-100 units