Sonicaid Team Operator’s Manual
91
10 User Maintenance
WARNING: ALWAYS SWITCH OFF THE TEAM AND DISCONNECT THE AC
SUPPLY CABLE AND TRANSDUCERS BEFORE ATTEMPTING TO CARRY
OUT ANY CLEANING OR MAINTENANCE.
10.1 Cleaning and sterilisation
Cleaning, general
Wipe the instrument case, transducers, event marker, fetal ECG electrode leg plate
and IUP extension cable with a cloth dampened in soap or detergent solution to
remove aquasonic gel, blood, saline etc. Wipe dry with a clean cloth.
Disinfecting the maternal ECG lead
1 Wipe with a cloth soaked in a solution of chlorine bleach in water (no stronger
than 1:10 mixture) or in a 2% Glutaraldehyde solution, such as Cidex.
2 Wipe the lead with a clean damp cloth, then a clean dry cloth.
Caution: do not use isopropyl alcohol. Do not expose metal components
(eg snap connectors) to chemicals.
Disinfection, general
Clean the instrument case, transducers etc. as described above. Then wipe with an
alcohol-impregnated wipe (70% ethanol or isopropranol).
Sterilisation
The only method of sterilisation for case and transducers is by using Ethylene Oxide
gas (up to 5.5 bar). Low-temperature steam is NOT permissible.
Note: sterilisation is not normally required.
Transducer care
Transducers should be kept dry and preferably below 45°C. Gel must be wiped from
the ultrasound transducers after use, and before placing on the storage area on the
side panel.