Sonicaid Team Operator’s Manual
90
9.3 Problems replaying or printing traces
Problem Remedy
On Team Duo I can’t hear the
1.5MHz transducer.
Use the loudspeaker enable button on the
MENU (left-hand side of the display).
Problem Remedy
When I try and reprint a stored
trace, the CTG trace is blank.
On Team Duo and Team IP you must select
which transducer you wish to store. It may be
that you are using the yellow transducer, but
have selected the blue transducer for storage.
Problem Remedy
The Toco channel of the CTG
graph is ‘scratchy’ and distorted.
I have tried using a different
Toco transducer, but it displays
the same symptom.
You may have enabled Actogram and be
unaware of the effect it has on the trace. Try
turning Actogram off in the MENU.
9.4 Team cycling from Logo screen to off
TEAM is ‘cycling’ from the Logo
screen to off.
The microprocessor can (rarely) get confused.
To resolve the problem:
1 Remove one end of the battery.
2 Short out the battery contacts.
3 Resolder the battery back again.
The unit will then function properly.