SLIT LAMP HS
-
7
5
00
41
11.5.
Immunity and Compliance Level
Immunity test
IEC 60601 Test
Level
Actual Immunity
Level
Compliance
Level
Conducted RF
IEC 61000
-
4
-
6
3Vrms
150kHz to 80MHz
3Vrms
3Vrms
Radiated RF
IEC 61000
-
4
-
3
3Vrms
80MHz to 2.5GHz
3V/m
3V/m