CareSens Control Solution:
You may check your meter and test strips using the
CareSens Control Solution.
The CareSens Control Solution contains a known amount
of glucose and is used to check that the meter and the
test strips are working properly.
The test strip vials have CareSens Control Solution ranges
printed on their labels. Compare the result displayed on
the meter to the CareSens Control Solution range printed
on the test strip vial.
Before using a new meter or a new vial of test strips, conduct a control
solution test following the procedure on page 23.
Note:
Use only the CareSens Control Solution.•
Check the expiration dates printed on the vial. When you first •
open a control solution vial, record the discard date (date opened
plus three months) in the space provided on the label.
Make sure your meter, test strips, and control solution are at room •
temperature before testing. Control Solution tests must be done
at room temperature (20 ~ 25˚C, 68 ~ 77˚F).
Before using the control solution, shake the vial, discard the first •
few drops and wipe the tip clean.
Close the control solution vial tightly and store at a temperature •
of 8 ~ 30˚C (46 ~ 86˚F).
Checking the System
Do a control solution test:
When using the meter for the first time.•
Whenever you open a new vial of test strips.•
If the meter or test strips do not function properly.•
If your symptoms are inconsistent with the blood glucose test results. •
and you feel that the meter or test strips are not working properly.
If you drop or damage the meter.•
[3] Press the M or the C Button to select the
correct code number. Each time you press
the M button, the number will increase by
one, and each time you press the C Button,
the number will decrease by one. Press and
hold the C Button to scroll code numbers
down.
[4] After selecting correct code number,
wait for three (3) seconds until the
symbol appears on display.
[5] When the symbol appears,
meter is ready for testing.