GUIDANCE AND MANUFACTURER’S DECLARATION-ELECTROMAGNETIC IMMUNITY
Icare PRO (TA03) is intended for use in the electromagnetic enviroment specied below. The customers
or users of Icare PRO (TA03) should assurre that it is used in such enviroment.
Immunity test IEC 60601 Testlevel Compliance
level
Electromagnetic environment-
Guidance
Voltage dips, short
interruption and
voltage variations on
power supply lines
IEC 61000-4-11
<5 % UT (>95% dip
in UT) for 0,5cycle
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT (30 % dip
in UT) for 25cycles
<5 % UT (>95% dip
in UT) for 5s
NOT
APPLICABLE
Icare PRO (TA03) tonometer is not
operational when connected to an
external power supply
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Radiated RF
IEC 61000-4-3
3 V/m
80MHz to 2.5 GHz
3V/m Portable and mobile RF
communications equipment should
be used no closer to any part of the
Icare PRO (TA03), including cables,
than the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Conducted RF
IEC 61000-4-6
3Vrms
150 kHz to 80MHz
NOT
APPLICABLE
Recommended separation
distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2 5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey should
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol: