Id300 Short-Pulse Laser Source Operating guide
8
4 Maintenance
4.1 Transportation and Storage
Maintain a temperature range within specifications when transporting or storing the
unit. Transportation damage can occur from improper handling. The following steps
are recommended to minimize the possibility of damage:
Pack the unit in the original packing material when shipping.
Store unit at room temperature in a clean and dry area.
Avoid high humidity or large temperature fluctuations.
Keep the unit out of direct sunlight.
Avoid unnecessary shock and vibration.
4.1.1 Safety Precautions
While manipulating optical fibres, laser radiation may be encountered at source output
ports and at fibre ends. Avoid long-term exposure to laser radiation.
Warning
Never look directly into a live optical fibre and ensure that your eyes are
protected at all times.
The following safety precautions must be observed during the operation and servicing
of the unit. Failure to comply with these precautions or with specific indications
elsewhere in this manual violates safety standards of intended use of the unit. id
Quantique assumes no liability for the user's failure to comply with these requirements.
This unit is intended for indoor use only.
Unit covers must not be removed at all time.
Before powering on the unit, all grounding terminals, extension cords, and
devices connected to it should be connected to a protective ground via a
ground socket. Any interruption of the protective grounding is a potential shock
hazard and may cause personal injury.
Whenever the ground protection is impaired, the unit is not to be used and
must be secured against any accidental or unintended operation.
Only fuses with the required rated currents and specified type (T1.6A) may be
used for replacement. Do not use repaired fuses or short-circuited fuse
holders.
Any adjustments, maintenance, and repair of opened unit under voltage
should be avoided and carried out only by skilled personnel aware of the
hazards involved. Do not attempt internal service or adjustment unless another
person qualified in first aid is present. Do not replace any components while
power cable is connected.
Operation of any electrical instrument around flammable gases or fumes
constitutes a major safety hazard.
Installation of replacement parts or modification of the unit must be carried out
by id Quantique. There is no service allowed by the customer.
The unit is to be returned to id Quantique for any service or repair.
Certain components inside the unit – e.g. capacitors may be charged even if
the unit has been disconnected from its electrical supply.