Mode of Operation
▪ For continuous use with short time loading.
▪ Duty Cycle: 10% (2 min on/18 min off).
Type of Equipment:
▪ Class II Type B applied part (as defined by IEC 60601-1, UL 60601-1,
EN 60601-1, CAN/CSA 601.1-M90, IEC 60601-2-46:1998.
▪ Type B protection against electrical shock as the applied part is the table surface.
Electrical:
▪ Supply Voltage: 120±5% Vac 60Hz or 2305% Vac 50Hz.
▪ Duty Cycle: 10% (2 min on/18 min off).
▪ Current Rating: Less than 10 Amps.
▪ Battery Backup Power.
Environmental:
▪ Operating Temperature Range: -10C to +40C.
▪ Operating Humidity Range: 30% to 75% relative humidity, noncondensing.
▪ Operating Pressure Range: 700 hPa to 1060 hPa.
▪ Transport & Storage Temperature Range: -40C to +60C.
▪ Transport & Storage Humidity Range: 30% to 75% relative humidity, noncondensing.
▪ Transport & Storage Pressure Range: 500 hPa to 1060 hPa.
▪ Rated IPX4 (Protected against splashing water).
▪ Meets EMC requirements of IEC 60601-1-2:2007.
Tabletop:
▪ The tabletop is made of carbon fiber and meets all the requirements of FDA CFR Title
21, Chapter 1, Subchapter J.