Table 3
This equipment is intended for use in the electromagnetic environment specified below.
The customer or user of this equipment should ensure that it is used in such an
environment.
Portable and mobile
communications equipment
should be separated from this
equipment by no less than the
distances calculated/listed below:
D=(3.5/3)(Sqrt P)
D=(3.5/3)(Sqrt P)
80 to 800 MHz
D=(7/3)(Sqrt P)
800 MHz to 2.7 GHz
where P is the max power in watts
and D is the recommended
separation distance in meters.
Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey,
should be less than the compliance
levels (V1 and E1).
Interference may occur in the
vicinity of equipment containing a
transmitter.