Operator Manual
Aspect 100-4T and ISR
L100-2843 Rev. E www.imagediagnostics.com 36
Electromagnetic
Environment
Guidance
±8kV Contact
±2kV, ±4kV, ±8kV,
±15kV Air
±8kV Contact
±2kV, ±4kV, ±8kV,
±15kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are synthetic, the
relative humidity should be
at least 30%.
±2kV at 100 kHz
repetition frequency
for AC Mains
±1kV at 100kHz
repetition frequency
for Signal I/O parts Port
Main power quality should
be that of a typical
commercial or hospital
environment.
±0.5kV, ±1kV Line to
Line
±0.5kV, ±1kV, ±2kV
Line to Ground
±0.5 kV, ±1 kV Line
to Line
±0.5kV, ±1kV, ±2 kV
Line to Ground
Main power quality should
be that of a typical
commercial or hospital
environment.
Voltage
Dips/Dropout
EN/IEC 61000-4-11
0 % UT; 0.5 cycle At 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % UT; 1 cycle
70 % UT; 25/30 cycles
for 50 Hz and 60Hz,
respectively
Single phase: at 0°
0 % UT; 250/300 cycle
for 50 Hz and 60 Hz,
respectively
Single phase: at 0°
0 % UT; 0.5 cycle At
0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle
70 % UT; 25/30 cycles
for 50 Hz and 60Hz,
respectively
Single phase: at 0°
0 % UT; 250/300
cycle for 50 Hz and 60
Hz, respectively
Single phase: at 0°
Main power quality should
be that of a typical
commercial or hospital
environment. If the user of
this equipment requires
continued operation during
power mains interruptions, it
is recommended that this
equipment be powered from
an uninterruptible power
supply or battery.
Power Frequency
IEC 61000-4-8
Magnetic Field
EN/IEC 61000-4-8
Power frequency magnetic
fields should be that of a
typical commercial or
hospital environment.
Guidance and Manufacturer’s Declaration- Immunity, All Equipment and Systems:
Table 2
This equipment is intended for use in the electromagnetic environment specified below.
The customer or user of this equipment should ensure that it is used in such an
environment.