20
Using wet cloths to remove dirt
Do not attempt to clean the reusable
cu󰀨 with water and never immerse the
cu󰀨 in water.
Disinfection: If required (for example, in a hospital setting), disinfection of
the pressure cu󰀨 twice per week is recommended.
Disinfect the interior of the pressure cu󰀨 (the surfaces that come into con-
tact with the patient’s skin) with the use of a soft cloth soaked in ethyl
alcohol (75-90%) and wait for the pressure cu󰀨 to dry up.
Cleaning of the pressure cu󰀨 after 200 uses is recommended.
Periodic calibration of device: the accuracy of measurement instru-
ments must be veried from time to time. For this reason, verication of
the indication of statistical pressure periodically (once a year) and after
technical maintenance and repair is recommended. More information can
be obtained from the service center.
SYMBOL INFORMATION
SYMBOL REFERENT
Consult instruction for use
Manufacturer
Manufacturing date