Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specied below.
The customer or the user of the device should assure that it is used in such an
environment.
Emissions test Compliance Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1 The device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B The device is suitable for use in all
establishments, other than domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/
icker emissions
IEC 61000-3-3
Complies
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specied below.
The customer or the user of the device should assure that it is used in such an
environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical
fast tran-
sient/burst
IEC 61000-
4-4
power supply
lines: ±2 kV
input/output
lines: ±1 kV
power supply
lines: ±2 kV
Mains power quality
should be
that of a typical com-
mercial or
hospital environment.