27
EN
Conducted
RF
IEC
61000-4-6
150 kHz to
80 MHz:
3 Vrms
6Vrms (in ISM
and amateur
radio bands)
80% Am at 1kHz
150 kHz to
80 MHz:
3 Vrms
6Vrms (in
ISM
and ama-
teur
radio bands)
80% Am at
1kHz
Portable and mobile RF com-
munications equipment should
be used no closer to any part
of the device, including cables,
than the recommended separa-
tion distance calculated from
the equation appropriate for the
frequency of the transmitter.
Recommended separation
distances:
d=0.35
P
;
d=1.2
P
Radiated
RF
IEC
61000-4-3
10V/m, 80% Am
at 1kHz
10V/m, 80%
Am
at 1kHz
80 MHz to
800 MHz:
d=1.2
P
800 MHz
to 2.7
GHz:
d=2.3
P
where, P is the
maximum output
power rating of
the transmitter in
watts (W)
according to
the transmitter
manufacturer,
d is the recom-
mended separa-
tion distance
in meters (m).
Field strengths
from xed RF
transmitters, as
determined by
an electromag-
netic site survey,
should be less
than the compli-
ance level in
each frequency
range.
Interference may
occur in the vicin-
ity of equipment
marked with the
following
symbol: