sization or irritation reaction.
* If you experience discomfort during a measurement, such as pain in the
arm or other complaints, press the START/STOP button to release the air
immediately from the cu󰀨. Loosen the cu󰀨 and remove it from your arm.
* If the cu󰀨 pressure reaches 40 kPa (300 mmHg), the unit will automati-
cally deate. Should the cu󰀨 not deate when pressures reaches 40 kPa
(300 mmHg), detach the cu󰀨 from the arm and press the START/STOP
button to stop ination.
* Before use, make sure the device functions safely and is in proper work-
ing condition. Check the device, do not use the device if it is damaged in
any way. The continuous use of a damaged unit may cause injury, im-
proper results, or serious danger.
* Do not wash the cu󰀨 in a washing machine or dishwasher!
* The service life of the cu󰀨 may vary by the frequency of washing, skin
condition, and storage state. The typical service life is 10000 times.
* It is recommended that the performance should be checked every 2
years and after
maintenance and repair, by retesting at least the requirements in limits of
the error of the cu󰀨 pressure indication and air leakage (testing at least at
50mmHg and 200mmHg).
* Please dispose of ACCESSORIES, detachable parts, and the ME
EQUIPMENT according to the local guidelines.
* Manufacturer will make available on request circuit diagrams, compo-
nent part lists,
descriptions, calibration instructions,etc., to assist to service personnel in
parts repair.
* The operator shall not touch output of batteries and the patient simul-
taneously.
* Cleaning :Dust environment may a󰀨ect the performance of the unit. Please
use the soft cloth to clean the whole unit before and after use. Don’t use any
abrasive or volatile cleaners.
* The device doesn’t need to be calibrated within two years of reliable
service.
* If you have any problems with this device, such as setting up, main-
taining or using, please contact the SERVICE PERSONNEL of Transtek.
Don’t open or repair the device by yourself in the event of malfunctions.
The device must only be serviced, repaired and opened by individuals at