EasyManua.ls Logo

IKA ULTRA-TURRAX T 25 basic - English; Safety Instructions

IKA ULTRA-TURRAX T 25 basic
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page
Safety instructions 17
Proper use 18
Unpacking 18
Useful information 18
Drive unit 18
Assembly of boom and stand 19
Working with the unit 19
Dispersion tools: General notes 19
Dispersion tool S 25 N - 20
Dispersion tool S 25 NK - 20
Dispersion tool S 25 KR - 21
Dispersion tool S 25 KV - 21
Dispersion tool S 25 KV -..- IL 22
Dispersion tool S 25 N - 8G 23
Dispersion tool S 25 N - 10G 24
Dispersion tool S 25 N - 10G - VS 24
Maintenance and cleaning 25
Accessories 25
Applicable standards and regulations 25
Guarantee 25
Technical data drive unit 26
Technical data for dispersion tools and properties 27
Technical data for dispersion tools and sealing types 28
Spare parts Drive unit T25 basic 80
Spare parts diagram T25 basic 81
Observe all pertinent safety instructions and directives, as well as work-
safety and accident-prevention regulations for laboratory use. The unit
must not be operated without a dispersion tool.
17
T25 1096EU
Never run dispersion tools dry, as the gasket and bearings will be
destroyed
if the tools are not cooled by the medium. The gasketand bearings are
made of PTFE and rustproof steel; the following points should therefore be
noted:
Chemical reactions of PTFE occur in contact with molten or solute alkali
metals and alkaline earth metals, as well as with fine powders of metals in
groups 2 and 3 of the periodic system at temperatures above 300 °C - 400
°C. Only elementary fluorine, chlorotrifluoride and alkali metals attack it;
halogenated hydrocarbons have a reversible swelling effect.
(Source: Römpps Chemie-Lexikon and ”Ulmann”, Volume 19)
Vibrations may loosen the hand screw on the flange of the drive unit. For
your own safety check that the hand screw is secure and tighten it if
necessary.
Glass vessels must always be secured with a clamp to prevent them
spinning. When working with flasks elastic couplers must be used to
prevent fracture of the glass.
Switch on the drive only at the lowest speed setting. Then gradually
increase the speed to the desired level.
Only dispersion and agitating tools approved by IKA may be used!
Even in the case of repair work, the unit must only be opened by a
qualified engineer. The mains plug must be pulled before opening the unit.
The voltage-carrying components inside the unit may continue to carry a
voltage for some time after the mains plug is pulled.
Caution: Pull the mains plug before changing the dispersion heads!
Caution: The integral switch
(2) is only single-pole and is thus not suitable
for safely isolating the unit from the supply!
The ventilation slots on the drive must not be obstructed!
The device does not start up again after a current failure.
The T 25 basic may not be used to process inflammable or explosive
media.
Contens
Safety instructions

Table of Contents

Related product manuals