How to troubleshoot flow rate error on illumina MiSeqDx?
- JJoseph JohnsonJul 31, 2025
The flow rate is the speed in which fluids pass through the fluidics system (µl/min). It is measured before each run during the pre-run check. If the system is unable to measure the flow rate, pump a volume of reagent (MiSeqDx SBS Solution (PR2)) through the system before checking the flow rate again. Use the drop-down list or on-screen keyboard to enter the following information: - Solution: PR2 - Volume: 250 µl - Aspirate Rate: 2500 µl/min - Dispense Rate: 2500 µl/min Select Pump. When the pump step is complete, select Restart Check. If the error persists, set the volume to pump 500 µl MiSeqDx SBS Solution (PR2) and repeat the process 1 more time. Contact Illumina Technical Support if the second attempt does not resolve the error.