FOR IN VITRO DIAGNOSTIC USE
ILLUMINA PROPRIETARY
Document # 15038353 v04
March 2018
MiSeqDx
®
Instrument
Reference Guide

Questions and Answers

  • A
    Alfred McconnellJul 31, 2025
    How to troubleshoot flow rate error on illumina MiSeqDx?
    • J
      Joseph JohnsonJul 31, 2025
      The flow rate is the speed in which fluids pass through the fluidics system (µl/min). It is measured before each run during the pre-run check. If the system is unable to measure the flow rate, pump a volume of reagent (MiSeqDx SBS Solution (PR2)) through the system before checking the flow rate again. Use the drop-down list or on-screen keyboard to enter the following information: - Solution: PR2 - Volume: 250 µl - Aspirate Rate: 2500 µl/min - Dispense Rate: 2500 µl/min Select Pump. When the pump step is complete, select Restart Check. If the error persists, set the volume to pump 500 µl MiSeqDx SBS Solution (PR2) and repeat the process 1 more time. Contact Illumina Technical Support if the second attempt does not resolve the error.
  • M
    Mr. Troy RichardsonAug 5, 2025
    How to resolve flow rate measured issue in illumina Laboratory Equipment?
    • J
      Jason BrownAug 5, 2025
      The flow rate check screen opens. Use the drop-down list or on-screen keyboard to enter the following: • Solution: PR2 • Volume: 250 • Aspirate Rate: 2500 • Dispense Rate: 2500 Select Pump. If the error persists, set the volume to pump 500 µl MiSeqDx SBS Solution (PR2) and repeat the process. When fluids have been pumped, select Restart Check. If the flow check fails again, reseat the flow cell to make sure that flow is not interrupted due to misalignment. Inspect the flow cell gasket for lint or irregularities.
  • S
    Sean BridgesAug 9, 2025
    How to prevent reboots during a run on MiSeqDx Laboratory Equipment?
    • E
      Ethan ThorntonAug 9, 2025
      If the MiSeqDx Laboratory Equipment restarts during a run, it could be due to Windows Update automatically installing software updates. To prevent this, contact your local IT department to disable automatic updates to the Windows operating system.
  • E
    eileendanielsAug 13, 2025
    What to do if the sample sheet is not present for illumina MiSeqDx?
    • K
      Kelly MercadoAug 13, 2025
      If the sample sheet is not named with the reagent cartridge ID, the instrument cannot automatically locate it. Browse to the correct sample sheet for the run. If the sample sheet is correctly named, ensure it is in the default sample sheet folder specified in Run Options on the Home screen. Confirm that the file extension is *.csv. If missing, create a sample sheet and copy it to the specified locations. After locating the sample sheet, select Restart Check.
  • S
    Selena MillerAug 18, 2025
    What to do if free disk space is low on illumina MiSeqDx Laboratory Equipment?
    • N
      Nicholas WilliamsAug 18, 2025
      If disk space is low, a message will appear indicating the required space. Use the Manage Files feature to clear the necessary space from the instrument computer.
  • K
    kellychristopherAug 21, 2025
    What to do if primary analysis is not ready on illumina MiSeqDx?
    • F
      fmartinezAug 22, 2025
      Primary analysis from the previous run is not complete. The default time to allow primary analysis to complete is 1 hour, and a countdown appears on the screen. You can either wait 1 hour or select Terminate Analysis. Note that secondary analysis stops for any incomplete cycles.
  • S
    Stephanie GarciaAug 27, 2025
    What to do if network connection is not active on illumina MiSeqDx Laboratory Equipment?
    • K
      Katelyn SmithAug 27, 2025
      Make sure the network cable is properly connected to the instrument. If the connection doesn't restore, reboot the software using the Reboot option on the Manage Instrument screen. If the problem continues, shut down the instrument via the Manage Instrument screen, turn off the power, wait at least 60 seconds, then turn the instrument back on and restart the software.

Summary

Getting Started

Intended Use

Definition and purpose of the MiSeqDx sequencing instrument.

Components

Description of the MiSeqDx instrument's exterior components.

Start the MiSeqDx

Step-by-step instructions for powering on and initializing the MiSeqDx instrument.

Required Consumables

List of consumables needed for the MiSeqDx, including user-supplied items.

Storage and Handling

Specifications for transportation, storage, and operating conditions of the instrument.

Anti-Virus Software

Recommendations for antivirus software configuration to ensure instrument operation.

MiSeqDx Software

Illumina User Management Software

Software for managing user names, passwords, and access levels for authorized users.

Illumina Worklist Manager Software

Application for creating and editing sample sheets for MiSeqDx and analysis software.

MiSeq Operating Software (MOS)

Software that guides users through run setup, controls instrument operation, and monitors runs.

Real-Time Analysis (RTA)

Integrated primary analysis software for image processing and base calling.

MiSeq Reporter

Secondary analysis software for processing base calls and producing sample information.

Performing a Run

Introduction

Overview of the steps required to perform a run on the MiSeqDx instrument.

MiSeqDx Workflow

Step-by-step process from sample library preparation to analysis.

Prepare the Reagent Cartridge

Instructions for thawing, inspecting, and loading the reagent cartridge.

Log In and Select Run Type

Steps to log into the software and select the appropriate run type.

Clean the Flow Cell

Procedure for cleaning the flow cell before loading it onto the instrument.

Load the Flow Cell

Instructions for correctly placing and securing the flow cell in the instrument.

Loading Reagents

Steps for loading the MiSeqDx SBS Solution (PR2) bottle and reagent cartridge.

Starting the Run

Reviewing run parameters and performing a pre-run check before initiating a run.

Monitoring the Run

How to track run progress, intensities, and quality scores on the Sequencing screen.

Maintenance Procedures

Perform a Standby Wash

Procedure to prepare the instrument for idle periods of 7 days or more.

Shut Down the Instrument

Steps to properly shut down the instrument software and hardware.

Troubleshooting

Troubleshoot Flow Rate Error

Steps to diagnose and resolve issues related to the instrument's fluidics flow rate.

Perform a Volume Test

Procedure to check fluidics system health by estimating volume between bubbles.

Resolve MiSeq Reporter Analysis Errors

Guidance for troubleshooting analysis errors, including re-queuing analysis.

Technical Assistance