5-E: Power Cord / Power Connection
Always make sure that the power switch is set to the OFF position before performing the
following tasks:
Plug the detachable AC cord into the back of the unit.
Plug the 3-prong plug into a grounded outlet.
A. Grounding:
Prior to connecting accessories to the unit, check that the main unit is plugged into a
grounded wall outlet.
B. Main voltage range and fuse:
Prior to plugging the AC adapter into the power outlet, check that the voltage is supported.
5-F: Foot Control Cable Assembly Connection
Align the pins of the foot control plug with the receptacle on the back of the device and push
in until the plug is firmly seated.
5-G: Water Supply Line Connection
Push the blue water tube into the stainless steel receptacle until the water tube cannot be
pushed any further. Then tighten the screw.
Connect the quick connect to the water supply line. The iM3 P6 should be prepackaged with
a male quick connect. If a female quick connect is necessary, please contact the iM3
representative to purchase the part.
Inspect all connections for leaks.
To remove the water line from the iM3 P6 scaler, first turn off the water supply or disconnect
the water supply line. Then loosen nut on the water tube from the receptacle of the unit and
gently pull the water tube out.
5-H: Hand piece connection and tip compatibility
The iM3 P6 hand piece is not only compatible with iM3’s tips, but also with tips designed for
Satelec Acteon piezoelectric scalers. The hand piece is also detachable from the hand piece
cable. During use, the operating frequency of the tip is measured to be approximately 29khz.