31
Step 1: First Command;
Write the X and Y values (Y only when learning a position) into registers 45891,
45892 (1702h, 1703h) and 45893, 45894 (1704h, 1705h) respectively.
Step 2: Second Command;
Write the “add to table” (010) command, or the “add learned table position” (101)
command to register 45889 (1700h).
** To read or delete a table position:
Step 1: First command;
Write the table position number of the position you wish to read or delete to register
45890 (1701h). (Table positions are numbered starting at 0 for the position with the
lowest X value. When a position is added or deleted from the table the table positions
are re-numbered.)
Step 2: Second command;
To delete the position from the table, write the “delete table position” (100) command
to register 45889 (1700h). Done
To read the table position data point pair, write the “load table position” (011)
command to register 45889 (1700h).
Step 3: Third command;
To read the table position data point pair, read the contents of registers 45891, 45892
(1702h, 1703h) for X and 45893, 45894 (1704h, 1705h) for Y.