If the pump system is operating, perform a sensitivity check
(see Sensitivity Check) on the sensor.
If your leak detector has failed to light 3 LEDs during the
sensitivity test, you should replace the sensor at the tip of probe
(see Installing the Sensor). With tip protector off, turn the leak
detector ON and look for a red glow in the small hole of the
sensor. If you fail to see a red glow, the leak detector has a
broken or shorted sensor power wire in the probe. If you see a
red glow, screw on the tip protector and perform the
Sensitiv-
ity Check
again. If your leak detector still fails to light 3 LEDs,
you have a broken or shorted sensor signal wire in the probe.
The above failures require you to take your leak detector to your
wholesaler for handling.
Problem
Your D-TEK will not power up (even after charging the proper time)
or all the lights illuminate with no audio sound.
Diagnostics/Remedy
This problem can be caused by power cells that need to be
recharged, improperly installed power cells, power cells that have
shorted or a faulty AC power adapter.
Note:
A new D-TEK Refrigerant Leak Detector does not come with the
batteries charged. To charge them see
Charging the Power Cells.
Note:
Power cells must be installed in the detector for proper operation.
To identify what area may be the problem you should:
Insure the power cells have been charged the proper amount of
time.
Note:
If the detector has been plugged into a power
source the proper amount of time, you may want to guarantee the
power source does not get turned off at night.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon D-TEK and is the answer not in the manual?

Inficon D-TEK Specifications

General IconGeneral
Power SupplyRechargeable battery
OutputAudible alarm
HousingPlastic
Battery Life8 hours, continuous
Operating Temperature32°F to 122°F (0°C to 50°C)