20
TROUBLESHOOTING GUIDE
SYMPTOMS
POSSIBLE FAULTS
POSSIBLE SOLUTIONS
Bed idle when hand pendant
buttons are pressed.
Adjustment may be at the
maximum or minimum
position
Power cord not connected or
damaged
Hand pendant or actuator
connectors are loose
Load is too heavy
Control system requiring a
reset
Check if other buttons are
working.
Ensure the power cord is
properly connected to the
electrical socket and the bed.
Ensure tight connection of all
connectors to control box.
Ensure the load weight is
within the designated
capacity. Reduce the load.
*Reset the electronic system
using the hand pendant.
Casters/Brakes noisy or stiff.
Debris or fluff in bearings
Replace casters.
Noisy or dry sound from
pivots.
Needs lubrication
Lubricate your bed.
Unusual noise from actuator.
Actuator is worn or damaged
or spindle is bent
Replace the actuator. Contact
your supplier.
Head/footboard unstable.
Head/footboard frame not
tightly secured.
Refer to Equipment Assembly
* Refer to page 18 of this the manual for electronics reset procedures.