3.1.7.1. The following information is a list of recorded parameters that can be
used to evaluate the device and its usage:
Real Time Clock
Life Clock
Flow Setting
Breaths per Minute
Low, Medium, and High Priority Errors
Ambient Pressure
System Temperature
Battery Status
Charging Status
3.1.8. Data Download
To download data from the Inogen One G3, you will need a USB mass storage device. If
one of the following drives cannot be obtained, the drive must be less than 2GB in
capacity and have a native file format of FAT32. The capacity of the drive and the file
system alone do not ensure compatibility with the Inogen One G3.
Compatible USB drives include:
PNY Attache 2GB PN: P-FD2GBATT2-SF
SanDisk Cruzer Micro 2GB PN:SDCZ6-2048RB
Kingston Data Traveler G3 2GB PN:DTIG3/2GBZ
Kingston DataTraveler 102 2GB PN:DT102/2GB
Kingston DataTraveler DT101 2GB PN: DT101G2/2GB
Dane-Elec DNL 2GB PN:DNLDAZMP02GCAW
Transcend JetFlash V30 2GB PN:TS2GJFV30
Patriot Signature 2GB PN: PSF2GUSB
To collect data:
a) With the Inogen One G3 off, insert the USB mass storage device into the USB port
behind the blue rubber cover
b) Plug in external power to the concentrator