INOGEN ONE G3 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
volume, and other physiologic factors. Inogen suggests that each patient be titrated
(a) while sedentary, (b) while active or ambulatory.
2.4.4. Use with the Inogen Satellite Conserver
The Inogen One G3 is not compatible with the Inogen One Satellite Conserver
3. SERVICING THE INOGEN ONE G3
3.1. Maintenance by the Provider
3.1.1. Authorized Repair Centers
The Inogen One G3 is only intended to be repaired by authorized repair centers. If a
repair is required, please contact Inogen to locate your nearest authorized repair
center. If you would like to become an authorized repair center, please contact Inogen
for access to certification training, service instructions, component part lists and the
necessary repair equipment.
3.1.2. Checking Life Clock, Serial #, and SW version
To check usage, press and hold the Alarm Bell Button for 5 seconds. The
display will show the hour meter
installed on the device.
Do not disassemble the Inogen One G3 or any of the accessories or attempt any
maintenance other than tasks described in this Technical Manual unless you have
completed a training course through Inogen. Disassembly of the Inogen One G3
or any of the accessories without proper training certification will void the product
warranties. Contact Inogen for information about receiving proper training and
certification for service of the Inogen One G3 and accessories.
3.1.3. Suggested Materials for Regular Maintenance
To perform regular field maintenance (by a technician) on the Inogen One G3 System,
you may need: