INOGEN ONE G3 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 16 of 20
or its life cycle and/or capacity will be greatly diminished. A full battery can be
damaged in as little as 20 days if left in the concentrator while not plugged in.
Instruct users to avoid storing the Inogen One G3 battery in extreme
temperatures, below -4
o
F (-20
o
C) or above 140
o
F (60
o
C), for any amount of time.
They should avoid leaving batteries in automobiles, where these temperatures
can be regularly exceeded. Storage of the Inogen One G3 battery in a cool, dry
location will help to maximize the longevity of the battery.
3.3. Expected Service Requirements
The following table is provided as an estimate only and assumes typical environmental
conditions for temperature, humidity, and air pollution. Any smoking around the
device will severely shorten its life expectancy. Please refer to product warranty
coverage terms.
Service Item
Estimated Service Requirement Frequency
Battery
500 full charge/discharge cycles (to approximately 80% capacity)
Compressor
Function of flow setting and time
20,000 run hours @ flow setting 4
25,000 run hours @ flow setting 2
Oxygen Sensor
No Service Requirement
Molecular Sieve
18 to 36 months depending on usage
Valves
25,000 run hours
Intake Filter
20,000 run hours (without smoke or other abnormal contaminants)
Cooling Fan
25,000 run hours
4. INOGEN ONE G3 SYSTEM SPECIFICATIONS
INOGEN ONE G3 CONCENTRATOR
Dimensions:
Length
Width
Height