70
Determination of aerosol output
2)
Fill the nebuliser with 2 ml of a 1% test solution.
Attach the nebuliser to the Breath Simulator until the aerosol
becomes irregular, plus one more minute.
Aerosol output in [ml] for minimum flow (3.0 l/min) and maximum
flow (6.0 l/min):
(Averages from each of 2 separate measurement series with 3 nebuliser/com-
pressed air combinations)
Determination of the aerosol output rate
3)
Fill the nebuliser with 2 ml of a 1% test solution. Attach the nebu-
liser to the Breath Simulator for at least 1 minute.
Aerosol output rate in [ml/min] for minimum flow (3.0 l/min) and
maximum flow (6.0 l/min):
(Averages from each of 2 separate measurement series with 3 nebuliser/com-
pressed air combinations)
We reserve the right to make technical changes.
Information as of: 2019-05
2) Aerosol quantity output by the nebuliser system for a specified fill
volume
Min. flow (3.0 l/min) 0.21 ml
Max. flow (6.0 l/min) 0.19 ml
3) Quantity of aerosol that is output by the nebuliser system per unit of
time
Min. flow (3.0 l/min) 0.096 ml/min
Max. flow (6.0 l/min) 0.144 ml/min