Measurement:
1. Unit:
The measuring unit can be selected under this menu. Different
range models may have different unit selection capabilities.
2. Group:
When several test samples need to be measured, the samples
can be coded into groups. The range is 01-99.
3.Test Mode:
Test mode can be selected. There are 5 kinds of mode: Track,
Peak , Preset, First Peak and Double Peak.
Preset:
In the Tolerance menu, program upper and lower limit for
GO/NG Measurement.
The upper limit value must be greater than the lower limit, and
both limit value can not be greater than 110% of the rated
capacity.
FirstPeak/Double Peak:
With 5% increase or decrease in the drop ratio of the
adjustment amount setting.
FirstPeak/Double Peak:
With 5% increase or decrease in the drop ratio of the
adjustment amount setting.
2
Memory:
In this menu, browse the data in memory or delete it/them and
Print data.
1. Browse Data
The greatest number is the most recent data.
2. Delete data:
There are 2 methods to delete the data for restoring the memory.
Delete selected: Delete data in number range selected.
Delete All: Delete all data saved.
Before delete data, a warning window will pop up for further
confirmation.
3. printing:
In Printing menu, you can Print Recent, Print Selected and Print
All.