7
8
Press key to enter the main menu interface.
Press key or to move the cursor to “System setup” and
then press to enter it.
Press key or to move the cursor to “Scan Mode” and then
press to switch on or off scan mode as below.
Press key to confirm setting and return to main menu
interface.
2
Alarm mode setting(Limit Setting):
The Thickness Gauge will alarm when the measured value is
beyond measurement range. The buzzer will ring if the measured
value reaches pre-set lows or pre-set highs.
Enter alarm mode setting according to “ scan mode setting”.
Press to turn on the alarm mode and enter menu ”LOW
LIMIT” as below.
Press key or to change lower limit.
Press key to enter menu ”HIGH LIMIT” to set high limit.
Press key to confirm setting and return to main interface.
3
Difference mode setting:
Enter difference mode setting according to “ scan mode
setting”.
Press to turn on the difference mode and enter menu
“TARGET MENU” as below:
Press key or to change the target value and press key
to go back.
Press key to confirm setting and return to main interface.
When making measurement, if the measurement value is larger
than target value, mark is replaced by + and if the measurement
value is larger than target value, mark is replaced by -. The value
showing on the screen is absolute of measurement value minus
target value.
4
Two-point CAL setting:
This procedure requires that the operator has two known
thickness points on the test piece that are representative of the
range to be measured.
Enter two-point CAL setting according to “ scan mode setting”.
Press to turn on the two-point CAL then press key return
to main interface.
When thickness value shows on the screen, press key to
make 2-Point CAL, the tester remind you to make thin
calibration as below: