Power On
Power on the Generator by pushing the power buon. The power buon will illuminate blue. An error in the start-up
indicates a device defect. Refer to the Error Diagnosis/ Troubleshooting section of this IFU.
Self-Test
The Generator performs a self-test aer it is powered on. All display elements on the front panel are turned on during this process.
The following light up briefly in the order they are listed: The MONOCUT and BICUT mode buons, then MONOCOAG and
BICOAG mode buons, followed by the red indicator light above the neutral electrode receptacle. The two Output Power displays
light up with “88” and the white CUT and COAG Indicator lights illuminate sequentially; the respective activation tone must sound
simultaneously. The default start-up of the device is BICUT and BICOAG with both displays lit up with “00”.
An error in the self-test sequence indicates a device defect. Refer to the Error Diagnosis/ Troubleshooting section of this IFU.