EN – ENGLISH
15
Table 2
Guidance and manufacturer’s declaration
– electromagnetic immunity
The CERTAS Plus Electronic Tool Kit is intended for use in the
electromagnetic environment specified below. The customer or the user
of the Tool Kit should ensure that it is used in such an environment.
IMMUNITY test
IEC 60601
test level
Compliance level
Electromanetic
environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30 %.
Electrical fast
transient/burst
IEC 61000-4-4
Not applicable Not applicable Not applicable
Surge
IEC 61000-4-5
Not applicable Not applicable Not applicable
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Not applicable Not applicable Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.