EN – ENGLISH
16
Table 3
Guidance and manufacturer’s declaration
– electromagnetic immunity
The CERTAS Plus Electronic Tool Kit is intended for use in the
electromagnetic environment specified below. The customer or the user
of the Tool Kit should ensure that it is used in such an environment.
IMMUNITY test
IEC 60601
test level
Compliance
level
Electromanetic
environment – guidance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.7 GHz
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Tool Kit than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmier.
Recommended separation
distance
d
= 1.2 √
P
150 kHz to 80 MHz
d
= 1.2 √
P
80 MHz to 800 MHz
d
= 2.3 √
P
800 MHz to 2.5 GHz
where
P
is the maximum output
power rating of the transmier
in was (W) according to the
transmier manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmiers, as determined
by an electromagnetic site
survey,
a
should be less than
the compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.