4.2 Alarm tests (continued)
Console Testing to IT0032 Issue 2
With the console set up and operating in
the RUN state disconnect the cooling
reservoir from the console and check:
The warning triangle and cooling alarm
indicator and the alarm tone can he heard
after approximately 30 seconds
Reconnect the reservoir and check:
The warning alarms visual and audio
cancel after approximately 30 seconds and
the console functions normally
Paper towels will be required to mop up
cooling water returning to the reservoir
Console Full Function Testing