CUSA EXcel
4.2 Alarm tests (continued)
Console Testing to IT0032 Issue 2
With the console set up and operating
in the RUN state loosen the hand piece
tip and set the amplitude to 10% then
press the amplitude pedal on the
footswitch and check:
The warning triangle and the tip alarm
indicator is active and the alarm tone
can he heard
Retighten the tip and check:
The warning alarms visual and audio
cancel and the hand piece functions
normally
Console Full Function Testing