Console Testing to IT0032 Issue 2
6. Power and Frequency output measurements
ALL of the following test are carried out
using the Yokogawa WT210 or WT310
digital power meter connected to the power
meter test socket on the ultrasonic board
with a pre-made test lead and connector
The power meter displays are set to:
Watts, set to 2 decimal places
Frequency, set to 2 decimal places
Power Factor, set to 2 decimal places
(optional)
NO other power meter has been validated for use
The WT310 will require additional software added
The WT310E is NOT validated for use