Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The AFFINITY COMPACT is intended for use in the electromagnetic environment specified below. The
customer or the user of the AFFINITY COMPACT should assure that it is used in such an environment.
Electromagnetic
Environment-Guidance
Discharge (ESD)
IEC 61000-4-2
+15 kV air
+15 kV air
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be greater
transient/burst
IEC61000-4-4
lines
+1 kV for input/output
lines
supply lines
+1 kV for input/output
lines
Mains power quality should be
that of a typical commercial or
residential environment.
IEC 61000-4-5
+2 kV common mode
mode
+2 kV common mode
Mains power quality should be
that of a typical commercial or
residential environment.
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
(>95% dip in UT) for 0.5
cycle
40% UT
(60% dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles
<5% UT
(>95% dip in UT) for 5
UT)
for 0.5 cycle
40% UT (60% dip in
UT) for
5 cycles
70% UT (30% dip in
UT) for
25 cycles
<5% UT
Mains power quality should be
that of a typical commercial or
residential environment. If the
user of the AFFINITY
COMPACT requires continued
operation during power mains
interruptions, it is
recommended that the
AFFINITY COMPACT be
powered from an
uninterruptable power supply
or its battery.
(50/60 Hz)
IEC 61000-4-8
fields should be at levels
characteristic of a typical
location in a typical
commercial or residential
Note: UT is the A.C. mains voltage prior to application of the test level.