D-0004578-R 2022/05 Page 36
Affinity2.0/Equinox2.0 - Instructions for Use - US
On Top Mode button converts the REM440 into an on top window that
includes only the most essential REM f eatures. The window is
automatically placed on top of other active sof tware programs such as the
relevant hearing aid f itting software.
When adjusting the gain handles in the fitting software, the REM440
screen will stay on top of the hearing aid f itting screen, allowing f or easy
curve comparison.
To return to the original REM440 press the red cross in the upper right-
hand corner.
Tube calibration button activates the tube calibration. Bef ore
measuring it is recommended to calibrate the probe tube. This is done
by pressing the calibration button. Follow the instructions appearing on
the screen (see screen below) and press OK. The calibration will then
automatically be perf ormed resulting in the curve below. Note that the
calibration is sensitive to noise and the clinician should theref ore ensure
that the room is quite while calibrating.

Other manuals for Interacoustics Affinity2.0

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics Affinity2.0 and is the answer not in the manual?

Interacoustics Affinity2.0 Specifications

General IconGeneral
BrandInteracoustics
ModelAffinity2.0
CategoryHearing Aid
LanguageEnglish

Related product manuals