SECTION 4—MAINTENANCE
Part No 1150695 29 Recessed Side Entry Bath Tubs
SECTION 4—MAINTENANCE
NOTE:Everysixmonthsorasnecessaryhaveaqualifiedtechnicianperformathoroughinspection
andservicing.Routinemaintenancewillextendthelifeandefficiencyofyourtub.
Safety Inspection Checklists
Initially
Ensurethatthemaximumwatersupplytemperaturedoesnotexceed115°F.
Ensurethatthemaximumwaterpressuredoesnotexceed65PSI.
Verifythewaterflowtemperaturewithaseparatethermometer.
Visuallyinspectforcracksinthetubsurface.
Ensurethatthedoorhandlelocksthedoorclosed.
Ensurethatthesafetypinproperlyengagesthedoor.
Checkthatalllabelsarepresentandlegible.Replaceifnecessary.
Before Each Use
Disinfectthetub.
Ensurethatthemaximumwatersupplytemperaturedoesnotexceed115°F.
Ensurethatthemaximumwaterpressuredoesnotexceed65PSI.
Visuallyinspectforcracksinthetubsurface.
Ensurethatthedoorhandlelocksthedoorclosed.
Ensurethatthesafetypinproperlyengages
thedoor.
Every 80-150 Baths
Performtheheavydutycleaningprocedureonthetub.RefertoHeavyDutyCleaning
Procedureonpage 31.
Every Six Months
Verifythewaterflowtemperaturewithaseparatethermometer.
Checkthatalllabelsarepresentandlegible.Replaceifnecessary.
CleanRADAvalvetoremovelimedeposits.