3
Use only original accessories supplied by Inventis srl, unless
specifically instructed otherwise.
Only use power adapters intended for medical equipment,
certified to IEC 60601-1. For further information, see
APPENDIX A: Technical Specifications”.
The Piccolo audiometer is a medical device. Any other external
device to which it is connected (such as a computer or CD player)
within the “patient area” (as defined in IEC 60601-1) must also
be a medical device, or must be protected by an isolating
transformer in order to ensure that the complete combination
(computer or external device + audiometer) complies with IEC
standard 60601-1.
Piccolo needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC
information provided in Appendix C.
Use of portable and mobile RF communications equipment can
affect the correct operation of Piccolo device. Make reference to
the EMC information in the Appendix C.
Calibration
The calibration of the audiometer is valid only for the transducers
supplied with the device. If a transducer is replaced, the
audiometer must be recalibrated.
The calibration of the audiometer is valid for the transducers
supplied with the audiometer only if these are connected directly
to the instrument without extension leads or other devices
between their connectors and the instrument (as habitually
occurs in soundproofed booth installations). If transducers are
not connected directly to the audiometer, the instrument must be
recalibrated before use.
Take note of the audiometer’s specified calibration interval. Use
of the instrument beyond its calibration expiry date can lead to
unreliable diagnoses.
In each test window, when you select a not-calibrated transducer,
the background of the ‘output’ area will be displayed in red color.
Moreover, you will not be able to send any stimulus through not-
calibrated transducers.

Other manuals for Inventis PICCOLO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inventis PICCOLO and is the answer not in the manual?

Inventis PICCOLO Specifications

General IconGeneral
BrandInventis
ModelPICCOLO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals