44 English
Power supply Battery powered mode: 2 x AAA batteries (3V DC)
Display mode Digital LCD V.A. 32 mm x 45 mm
Measurement mode Oscillographic testing mode
Measurement range
Rated cuff pressure: 0 mmHg~300 mmHg (0 kPa ~40 kPa)
Measurement pressure: SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40 mmHg~130 mmHg (5.3 kPa~17.3 kPa)
Pulse value: (40-199) beat/minute
specifications
13
Problem Symptom Check this Remedy
Error message
E6 shows
The voltage of the batteries is
too low during the inflation
Measure again. If E6 appears again, please replace with
new batteries.
E10 or E11 shows
The monitor detected motion
while measuring.
Movement can affect the measurement. Relax for a
moment and then measure again.
E20 shows
The measurement process
does not detect the pulse
signal.
Loosen the clothing on the wrist and then measure
again.
E21 shows Measure incorrectly. Relax for a moment and then measure again.
EExx shows on the display A calibration error occurred.
Retake the measurement. If the problem persists,
contact the retailer or our customer service for
further assistance. Refer to the warranty for contact
information and return instructions.