EasyManua.ls Logo

ISA DRTS.3 PLUS - 2 Applicable Standards; Electromagnetic Compatibility; Emission; Immunity

ISA DRTS.3 PLUS
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Doc. SIE62153 Rev. 1 Page 6/24
2 APPLICABLE STANDARDS
The test set DRTS.3 PLUS and optional modules conform to the EEC directives regarding
Electromagnetic Compatibility and Low Voltage instruments.
2.1 Electromagnetic Compatibility
Directive no. 2004/108/EC. Applicable Standard : EN61326-1 + A1 + A2.
EMISSION
- EN 61000-3-3: Limitation of voltage fluctuations and flicker. Acceptable limits: basic.
- CISPR16 (EN 55011 class A): Limits and measurement methods of radio-electric disturbances for
industrial, medical and scientific instruments at radio-electric frequencies.
Acceptable limits for conducted emission:
. 0.15-0.5 MHz: 79 dB pk; 66 dB avg.
. 0.5-5 MHz: 73 dB pk; 60 dB avg.
. 5-30 MHz: 73 dB pk; 60 dB avg.
Acceptable limits for radiated emission:
. 30-230 MHz: 40 dB (30 m)
. 230-1000 MHz: 47 dB (30 m)
IMMUNITY
- EN 61000-4-2: Immunity tests for ESD. Test values: 8 kV in air; 4 kV in contact.
- EN 61000-4-3; Immunity tests for radio frequency interference. Test values (f= 900 ± 5 MHz):
field 10 V/m, modulated AM 80%; 1 kHz
- EN 61000-4-4; Immunity tests for high speed transients (burst). Test values: 2 kV peak; 5/50 ns.
- EN 61000-4-5; Immunity tests for surge. Test values: 1 kV peak differential mode; 2 kV peak
common mode; 1.2/50 us.
- EN 61000-4-6: immunity to low-voltage sinusoidal waveform. Test values: 0.15-80 MHz, 3 Vrms,
80% AM 1 kHz.
- EN 61000-4-8: Immunity tests for low frequency magnetic fields. Test values: 30 Arms/m.
- EN 61000-4-11: Immunity test for power supply drops. Test value: 1 cycle; 100% drop.
1.2 Low voltage directive
Directive n. 2006/95/EC. Applicable standard: CEI EN 61010-1. In particular:
- Dielectric Rigidity: 1.4 kV, 1 minute.
- Isolation resistance: > 2 MOhm.
- Earth resistance : < 0.1 Ohm.
- Dispersion current: < 5 mA.
- Inputs/outputs protection: IP 2X - IEC 60529.
- Operating temperature: 0 - 50°C; storage: -25°C to 70°C.
- Relative humidity : 5 - 95%, not condensing.
- Vibration: IEC 68-2-6 (20 m/s^2 at 10 150 Hz);
- Shock: IEC 68-2-27 (15 g; 11 ms; half-sine).
- Altitude: less than 2000 m.