Function and Features 
  Copyright ©ITECH Electronic Co., Ltd.  14 
0:represents 1-10sets; 
1:represents:11-20sets,   
 
On(Def) 
enable the buzzer function
 
disable the buzzer function 
Select the communication interface
 
 
Select RS232 communication interface
 
4800,    8,    None parity 1 ,  Addr... 
19200     
  
Select USB communication interface
 
GPIB 
Select GPIB communication interface 
 
Addr= 0    set the communication address
 
RS485 
Select RS485 communication interface 
4800,    8,    None parity,    1  ,    Addr...
 
Enables automatic delay to switch display 
from setting to measured value (meter). 
 
Automatically Return Meter function 
disabled. 
The function of Automatically return meter 
after 5s enabled. 
Set the output status of power supply power 
on 
After power on, the power supply output 
status will be Off. 
The output status of power supply will be 
same as last output status before power off. 
Sense-Protect  Sense Protect   
Enable sense protection function 
 
Enable 
Disable sense protection function 
Disable(Def) 
Enables dummy load to increase speed of 
voltage fall time.