Remote Control 
  Copyright © Itech Electronic Co., Ltd.  8 
always resets current SCPI command path to root level. 
NOTE 
As for every SCPI message with one query sent to the instrument, the 
instrument will use a <NL> or newline sign (EOI) to terminate response of 
return. For example, if "DISP:TEXT?" is sent, <NL> will be placed after 
the  returned  data  string  to  terminate  response.  If  an  SCPI  message 
includes  several  queries  separated  by  semicolon  (such  as 
"DISP?;DISP:TEXT?"),  <NL>  will  terminate  response  returned  after 
response to the last query. In all cases, the program must read <NL> in 
response before another command is sent to the instrument, otherwise 
errors will be caused. 
1.7 Data Type 
SCPI  language  defines  several  data  types  used  for  program  message  and 
response messages. 
  Numerical parameter 
Commands requiring numerical Parameter support the notations of all common 
decimal notations, including optional signs, decimal points, scientific notation, 
etc. Special values of numerical Parameter are also acceptable, such as MIN, 
MAX  and  DEF.  In  addition,  suffixes  for  engineering  units  can  also  be  sent 
together with numerical Parameter (including M, k, m or u). If the command 
accepts only some specific values, the instrument will automatically round the 
input  Parameter  to  acceptable  values.  The  following  commands  require 
numerical Parameter of frequency value: 
[SOURce[1|2]:]FREQuency:CENTer {<Frequency>|MINimum|MAXimum} 
  <NR1>: There is an implicit decimal point in the last bit,such as 273 
  <NR2>: There is an explicit decimal point,such as .273 
  <NR3>: There are an explicit decimal point and exponential,such as 
2.73E+22.73E+2   
  <Nrf>:The extensible form includes <NR1>, <NR2> and <NR3>,such 
as 273 273. 2.73E2273 273. 2.73E2 
  <Nrf+>: The extensible decimal form includes <NRf> and MIN MAX 
DEF, such as 273 273. 2.73E2 MAX. MIN and MAX are the minimum 
and  maximum  finite  number.  Within  the  range  of  the  parameter 
definition, DEF is the default of the parameter. 
  Discrete parameter 
Discrete Parameter are used for settings with limited number of programming 
values (such as IMMediate, EXTernal or BUS). They can use short and long 
format like key words of commands. They may be expressed in both upper and 
lower case. The query response always returns uppercase Parameter in short 
format. The following commands require discrete Parameter in voltage unit: 
[SOURce[1|2]:]VOLTage:UNIT {VPP|VRMS|DBM} 
  Boolean parameter 
Boolean  Parameter refer  to  true  or  false  binary  conditions. In  case  of false 
conditions, the instrument will accept "OFF" or "0". In case of true conditions, 
the  instrument  will  accept  "ON"  or  "1".  In  query  of  Boolean  settings,  the 
instrument will always return "0" or "1". Boolean Parameter are required by the 
following commands: 
DISPlay {OFF|0|ON|1}